Disposable medical devices E-beam sterilization can be processed in different ways. It has been processed in carton boxes for many years. In this case it is often irradiated at an energy level of 10 MeV, although irradiation facilities work at lower energies.
Medical devices can also be treated piece by piece. This second case has the advantage to require lower energy level (cheaper accelerators and cheaper shielding). By the same token it requires one electron accelerator per product type.

 

1. Disposable medical devices in carton boxes |

 

What is the power requirement of a 10 MeV electron accelerator

 

 

that must sterilize (dose 25 kGy) the following boxes in

2000

hours

Considering that

 

 

 

 

 

1

Scanning Horn width is variable

 

 

 

2

Scanning Horn width is

80

cm

 

 

 

 

 

 

 

 

 

 

Width (cm)

Length (cm)

Height (cm)

Weight (kg)

m3

 

Box 1

40

50

22

6

         12.800  

 

Box 2

40

60

40

17

         14.800  

 

Box 3

60

75

45

15

          6.400  

 

Box 4

60

70

25

11

         17.300  

 

Box 5

60

80

30

20

          9.500  

 

 

 

 

 

 

         60.800  

 

 

 

 

 

 

 

Number of boxes (# boxes) is equal to the number of cubic meters / volume of each box

 

 

 

 

 

 

 

 

# boxes

 

 

 

 

 

Box 1

     290.909  

 

 

 

 

 

Box 2

     154.167  

 

 

 

 

 

Box 3

      31.605  

 

 

 

 

 

Box 4

     164.762  

 

 

 

 

 

Box 5

      65.972  

 

 

 

 

 

 

 

 

 

 

 

 

The Ratio (g/cm2) is Weight / (Width x Length) for each type of box
If the Ratio is > 3,6, the box requires double side irradiation.
The total surface to irradiate - practically - if the scan width is fixed is

 (width of scan  x  # boxes  x  Length)

 multiplied by 2 if this box requires double side irradiation
The total surface to irradiate - practically - if the scan width is variable

 if we consider 5cm over-scanning on each side

 ((5 + width of box + 5)  x  # boxes  x  Length)

 multiplied by 2 if this box requires double side irradiation

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Tot Surf at

Tot Surf at

 

 

 

 

 

80

+ 5 / + 5

 

 

 

Ratio (g/cm2)

Irradiation

(m2)

(m2)

 

 

Box 1

3,00

S. Side Ir.

     116.364  

      72.727  

 

 

Box 2

7,08

Double S. Ir.

     148.000  

      92.500  

 

 

Box 3

3,33

S. Side Ir.

      18.963  

      16.593  

 

 

Box 4

2,62

S. Side Ir.

      92.267  

      80.733  

 

 

Box 5

4,17

Double S. Ir.

      84.444  

      73.889  

 

 

 

 

 

     460.038  

     336.442  

 

 

 

 

 

 

 

 

 

Formula

A / T = K F I / D

 

 

 

 

 

 

 

 

 

 

 

1. Scan Width Variable (+5 cm on each side)

 

 

 

 

336.442

linear m at

variable

scan width in

2000,00

hours

 

This translates into

2,80

m2/min

 

 

A / T

2,80

m2/min

 

 

 

 

D

2,5

Mrad

 

I = A D / T K F

 

K

1,17

(Mrad m2/mA min)

I : Current =

7,49

mA

F

0,8

% efficiency

 

Power

74,88

kW

 

 

 

 

 

 

 

2. Scan Width = 80 cm

 

 

 

 

 

460.038

linear m at

80

scan width in

2000

hours

 

This translates into

3,83

m2/min

 

 

A / T

3,83

 

 

 

 

 

D

2,5

 

 

I = A D / T K F

 

K

1,17

 

 

I : Current =

10,24

mA

F

0,8

 

 

Power

102,3944334

kW

 